KM900PLUS • Combustion Analyzer
Copyright © 2007 UEi Specifications are subject to change without notice. L517 5/08
Distributed By:
1-800-547-5740 • Fax: (503) 643-6322
www
.ueitest.com • Email: [email protected]
IN CANADA
Vancouver: Phone: 1-877-475-0648 • (604) 278-4511 • Fax: (604) 278-8299
Toronto: Phone: 1-877-475-0648 • (905) 238-6760 • Fax: (905) 238-5117
IN EUR
OPE
Phone: +44 1707 375550 • Fax: +44 1707 393277
The KM900PLUS handheld analyser is easy to
use for domestic and industrial boiler flue gas
measurement and analysis. It operates on all boiler
types and can display 8 parameters simultaneously
showing measurements and calculations of boiler
efficiency and pollution.
The KM900PLUS measures O2, CO, flue and nett
temperature and calculates CO2, boiler efficiency,
excess air and losses. CO is measured with a
hydrogen compensated sensor for increased
accuracy.
The KM900PLUS
has the following features:
Measures O2 and CO
Calculates CO2, efficiency and excess air
0 to 10,000 ppm H2 compensated CO sensor
Large 4 line display
Long life rechargeable battery
Large cordura case
150 memory positions
One year limited warranty
Two year limited warranty on O2 sensor
Options:
up to 2 additional sensors
GAS RANGE PART#
NO1 0~5,000ppm 13251*
NO1 low range 0~100ppm 18079*
NO2 0~1,000ppm 10533
SO2 0~5,000ppm 13910
SO2 low range 0~100ppm 18131
*NOX will be calculated only if NO1 sensor is installed.
Specifications
PLEASE
RECYCLE
Parameter Resolution Accuracy Range
Temp. Measurement
Flue Measurement 1.0˚F/C ±2.0˚F/C, ±0.3% of rdg 32˚ - 1112˚F
with probe (0˚ - 600˚C)
Inlet Temperature 0.1˚F/C ±1.0˚F/C, ±0.3% of rdg 32˚ - 122˚F
(0˚ - 50˚C)
Pressure 0.1 mbar ±2% of full scale +150 mbar to -
(standard) 150 mbar
Gas Measurement*
1
Oxygen (02) 0.1% ±0.2% 0 - 21%
Carbon Monoxide (CO) 1 ppm ±20 ppm <400 ppm 0 - 10,000 ppm
(standard: H compensated) ±5% >5000 ppm
±10%<5000 ppm
Nitric Oxide (NO)
1 ppm ±5 ppm <100 ppm 0 - 5000 ppm
(optional) ±5%>100 ppm
Nitric Oxide (NO) 1 ppm ±2 ppm <30 ppm 0 ~ 100 ppm
(Low range)(optional) ±5 ppm <30 ppm
Nitrogen Dioxide 1 ppm ±5 ppm<100 ppm 0 ~ 1000 ppm
(optional) ±10 ppm<500 ppm
±5%>500 ppm
Sulphur Dioxide 1 ppm ±5 ppm<100 ppm 0 ~ 5000 ppm
(optional) ±5%>100 ppm
Calculations
Carbon Dioxide*2 0.1% ±0.3% reading 0 ~ 99.9%
Losses 0.1% ±1.0% reading 0 ~ 99.9%
Efficiency 0.1% ±1.0% reading 0 ~ 99.9%
Excess Air 0.1% ±0.2% 0 ~ 2885.0%
Temp (Nett) 1.0˚C/F ±2˚C ±0.3% reading 32 ~ 1112˚F
(0 ~ 600˚C)
CO/CO2 Ratio 0.0001 ±0.0001 0 ~ 0.9999
Poison Index 0.01% ±0.01 0 ~ 99.99
Parameter Description
Preprogrammed fuels Natural Gas, Town Gas, Gascor, Light Oil, Heavy Oil, Propane,
Butane, Anthrocite, Coke, Coal, Kinsale Gas
Dimensions: Weight 1 Kg
Handset 220 x 55 x 120 mm
Probe (L) 420 x (Dia) 8mm with stainless steel shaft, K-type
thermocouple and 3m hose
Ambient operating range 0˚C to 45˚C/10% to 90% RH non condensing
Power supply Input: 110V AC / 220V AC nominal
(Battery charger) Output: 12V AC off load
Battery life >6 hours from full charge