Be f o r e using this meter, read all safety information care f u l l y. In this manual the word "WARNING" is used
to indicate conditions or actions that may pose physical hazards to the user. The word "CAUTION" is used
to indicate conditions or actions that may damage this instru m e n t .
WA R N I N G !
This analyzer extracts combustion gases that may be toxic in re l a t i vely low concentrations. These gases are
exhausted from the back of the instrument. This instrument must only be used in we l l - ventilated locations.
It must only be used by trained and competent persons after due consideration of all the potential hazard s .
P
REFL IGH T
C
HECKL IST
Clean particle filter
Water trap and probe line are empty of water
Power on and ze r o
All hose and thermocouple connections are properly secure d
Flue gas probe is sampling ambient FRESH air
Water trap is fitted correctly to the instru m e n t
Flue temperature plug is connected
Inlet temperature probe is connected if re q u i r e d
A
NA LY Z ER
C
ONNEC T IONS
N OTE: Take care when inserting the temperature
p r obes as the pins are polarized. In s e rt with the
smaller pin (+) to the right. See diagram to right.
WA R N I N G !
Turning the pump off while the probe is in the flue
will leave toxic gases inside the analyze r. Once data
has been printed or copied it is advisable to purge the
unit with fresh air as soon as possible. To do this,
with the probe re m o ved from the flue, turn ON the
p u m p . Always allow the readings to return to ze r o
(20.9 for O2) prior to shutting the unit off. The
meter will not switch off until the CO reading is
b e l o w 20 ppm.
WA R N I N G !
The probe will be hot from flue gases. Re m o ve the
p r obe from the flue and allow it to cool naturally. Do
not immerse the probe in water, as this will be drawn
into the analyzer and damage the pump and sensors.
Once the probe is re m o ved from the flue and the
readings have returned to ambient levels hold down
“ O n / Of f ” and switch off the analyze r. The
i n s t rument will count down from 30 to switch off. If
you pressedOn / Off by mistake, pressingSend
will return you to normal operation.
P
O ST
F
L IGH T
Re m ove the probe from the flue and allow the analyzer to purge with fresh air until readings return to ze r o.
- O2 to 20.9%, CO to Ze r o (Be careful as the probe tip will be HOT )
Drain water trap
Check particle filter
OWNERS
MANUAL &
MAINTENANCE
Temperature Connections
• Flue Probe Temp: T1
• Inlet Temp: T2
Pressure Connections
• Single Input (Draft): P1
• Differential Input Testing
(Pressure drop test) P2
Flue Probe Temperature Plug
(Plugs into T1)
Narrow Pin MUST be on the
Right hand side.
Battery Charge/AC
Power Adapter
Flue Gas Inlet
Connection
Watertrap drain
Flue Probe Gas Inlet Plug
Setting Inlet Te m p e r a t u r e
- Turn on and ze r o the analyzer with
out the flue probe connected to use
ambient temperature
- Connect flue probe thermocouple to
T1 during ze r o countdown to store
p r obe tip temperature as inlet
(ducted system)